Users' Forum
EQUIPMENT

NANON

NANON
Electrospinning System for R & D

NANON is our best-selling electrospinning system for laboratory use. A wide variety of materials can be spun by programming the spinning conditions and combining system parts such as spinnerets and collectors.

400 Systems Delivered

Including 300+ NANON units, used at The University of Tokyo, MIT, University of Washington, King Saud University and many more.

See typical customers ›

Our specialist will reply by the next business day (Japan time).

Major features

  • Uniform, high-quality nanofibers
    MECC’s proprietary technology creates a stable electric field, producing nanofibers with a highly uniform diameter and very few droplets.
  • Easy programming with a 10-key pad
    Parameters such as high voltage, feed rate and traverse speed can be programmed easily.
  • Easy-to-change system parts
    Spinnerets, collectors and other parts can be changed and installed quickly and easily.
  • Aligned and core-shell nanofibers
    With a drum collector or a core-shell spinneret, you can spin special fibers such as aligned membranes and core-shell nanofibers.
  • Automatic cleaning mechanism
    The needle tips can be cleaned automatically at the frequency and intervals you program.
  • Thorough safety measures
    Door interlock, emergency stop, operation indicator lamp, tempered glass and high-speed high-voltage shutoff keep operators safe.

Product introduction video

Performance

* The effective width varies depending on the material.

Model name

NANON

Purpose of use Research and development
Typical nanofibers to spin Nonwovens of double size of A4 (200mm × 600mm)
Aligned membrane
Core/shell ・hollow nanofiber
Nanowire
Tubular fiber
Standard collector Plate
Optional collector Drum φ100 w200(100 to 3000rpm)
Drum φ200 w200(100 to 3000rpm)
Drum φ200 w30(100 to 3000rpm)
Disk(100 to 3000rpm)
Mandrel(50 to 100rpm)
Y-axis Slide (±25mm-wide)
Roll to roll 200mm-wide
Standard spinneret Clip(75mm)
Optional spinneret Vertical Clip
Tubeless
Coaxial
MJS 4 holes
Spinneret operation Range of motion 200mm
Tubeless
Coaxial
MJS 4 holes
Syringe pump 1 built-in(To mount 3 syringes)
HSP-101(External option)
Feed rate 0.1 to 100ml/hr
Control 10-key
High voltage power supply Voltage:0.5 to 30.0kV
Current:0 to 50.0μA
Spinning distance 50 to 150mm
Programmable parameter ・High voltage output
・Feed rate of solutions from pump
・Collector rotation speed
・Slider vertical travel (Semi-automatic)
・Frequency and interval of cleaning
・Slider speed and distance
Fan airflow (intake / exhaust) 12㎥/hr or more
Capacity of syringes 2.0 / 5.0 / 10.0 (ml)
External dimensions (Except protrusions) 830W x 615D x 875H (mm)
Weight 110kg or less
Safety devices ・Safety interlock
・Door lock
・Emergency stop button
・Signal tower
・Tempered glass
・Exhaust purification filter
・High speed shutoff of high-voltage output after POWER OFF
・High-voltage shutdown in case of overcurrent

Components to be integrated

Plate Collector

Drum Collector

Mandrel Collector

Disk Collector


Y-axis Slide Collector


Roll to Roll Collector

Clip Spinneret

Vertical Clip Spinneret

Coaxial Spinneret

Tubeless Spinneret


Multiple Jet Spinneret (MJS)


Syringe Pump

Talk to Our Electrospinning Specialists

Not sure if NANON fits your research? Tell us your material and goal.
We will help you choose the right setup.

Quote

Get a price for NANON with the spinnerets and collectors you need.

Online Meeting

Meet with us online to discuss your research and the right setup.

Sample Making

Consult us about making nanofiber samples to check the results before you buy.

Contact Us

Our specialist will reply by the next business day (Japan time).
You can also email us at askmecc@mecc.co.jp

Catalog / specifications

After filling out the form below, the product catalog will be automatically sent to the registered email address.


    RequiredOrganization



    URL



    RequiredName



    Required City



    Required Country



    RequiredPhone



    RequiredE-mail



    Comments/Requests